MY01 Base Station Rapid Resource Guide

S
Written by Sara Kassab
Updated over 2 weeks ago

MY01 Continuous Compartment Pressure Monitor

Step-by-Step Instructions

Note: If necessary, a small incision can be made to facilitate the insertion process.

Common User Errors

Error

Solution

Tip protector not removed before powering on

• If device zeros after protector removal, continue using

• If device fails to zero or pressures are not as expected, replace device

Inserted device before powering on

• If introducer is still attached and device has not been powered on, remove it, power on, re-insert

• If device is fully deployed and powered on inside the muscle, remove and replace device

Removed sensor with introducer

(did not rotate introducer 180°)

• Re-insert if the sensor is still hooked

• Replace device if the sensor is unhooked

Lead wire removed when adjusting device

• Secure lead wire with dressing immediately after insertion

• Replace device if the sensor is removed accidentally

Device sensor holes covered during use

• Expose sensors, continue monitoring

Unit Display Interface Symbols

Symbol

Title and Description

Bluetooth

Full battery

Warning: Low battery level

Trend Arrow: indicates rising or falling pressure values

Critical Errors - Discontinue Use

ERR 10: Sensor Compromised

ERR 205: Sensor Comunication Lost

ERR 403: Memory Corruption

ERR 11: Software Error

ERR 301: Display Unresponsive

ERR 404: System Resets

ERR 101: Maximum Zeroing Entries

ERR 401: System Overheating

ERR 405: Invalid System State

ERR 201 - 204: Sensor Tip Errors

ERR 402: System Too Cold

Non-Critical Errors

3: Lifetime Exceeded (>18 hours)

Use beyond this point is off-label

4: Outside Operating Pressure Range

The absolute pressure reading from sensors is outside operating pressure range. Disregard the pressure values while error sign is visible.

Reach out to MY01 customer support to facilitate return of unit for further diagnostic evaluation: info@my01.io or +1 (855)799-6901

MY01 Mobile Application

Understanding the MY01 Application Display

Connecting to the MY01 unit

  • If the unit does not immediately appear on the “active” tab, swipe down on the screen to refresh and scan for nearby MY01 units.

  • Once the unit appears, verify the device identifier matches on both the application and unit display.

  • Tap on the unit in the app to connect. If connecting for the first time, confirm pairing PIN matches PIN on unit display.

  • Bluetooth mark will appear on application and unit display when connected.

Evaluation Perfusion Pressure in the App

  1. Navigate to the data and trend display for the active unit.

    • Tap on “Diastolic Entry” in the upper left and input the patient’s diastolic blood pressure. Select date and time and confirm with checkmark (✓).

  2. A confirmation message will pop up.

    • Use “Copy to Device” if using multiple units to monitor one patient (ex. multi-limb evaluation)

    • Tap “Set New Reminder” to receive a notification for when you would like to update the diastolic entry again

    User-to-User Invite

  • Go to the Menu and tap "Invite Members"

  • New user should scan the QR code to join the organization.

  • Notes:

    • The QR code and link are only valid until the date at the bottom right.

    • If you are a part of multiple organizations, double-check the correct organization is selected (bottom left).

Need Help?

  • Use the menu link at the top left of the application to access the help center.

  • If you cannot find the answers in the help center, the chat link at the upper right of the app will connect you with a MY01 team member.

Did this answer your question?